Appendix B: Performance Verification
B-52 AWG610 Arbitrary Waveform Generator User Manual
1/4 Clock FrequencyĂandĂ10 MHz Reference Input TestsThese procedures check the 10 MHz reference input function of the AWG610
Arbitrary Waveform Generator.
Equipment
required
Two50ĂWBNCĂcoaxial cables, a frequency counter, and a function
generator.
Prerequisites The AWG610Arbitrary Waveform Generator must meet the
prerequisites listedon page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W BNC coaxial cable to connect the AWG610 Arbitrary Waveform
Generator 1/4 CLOCK OUT connector to the input A connector on the
frequency counter.
2. Use a 50 W BNC coaxial cable to connect the AWG610 Arbitrary Waveform
Generator 10 MHz REF IN connector to the function generator output
connector (see Figure B–28).
Function Generator (AFG310) Frequency Counter
AWG610Arbitrary Waveform
Generator rear panel
50 WBNCĂcoaxial cable
50 WBNCĂcoaxial cable
1/4ĂCLOCKĂOUT 10ĂMHzĂREFĂIN
Figure B-28: 1/4 Clock frequency and 10 MHzreference input initial test hookup
a. Set the frequency counter controls as follows:
INPUT A
Coupling . ................... AC
FUNCTION ...................... AFREQ
Gate time ....................... 0.2s
TriggerLevel . .................... 0V