The Sequence Editor
AWG610 Arbitrary Waveform Generator User Manual 3-117
Do the following steps to set the Wait Trigger value:
1. Move the cursor to the line in which to set the Wait Trigger value.
1. Move the cursor to the Wait Trigger column.
2. Push Data Entry (bottom).
3. Push Wait Trig. (side) to toggle between On and Off. The Off state is a
blank in the column.
4. Push the CLEAR MENU on the front panel to exit the setting mode.
The Goto One column lets you set an unconditional jump to the first line of the
sequence table (go to line one). Valid values are On and Off (blank). Goto One
functionality is only valid when the Run Mode is set to Enhanced. Note that this
parameter is ignored if it is set in a subsequence file.
The instrument processes sequence table entries until it encounters a Goto One.
If the instrument Run Mode is set to Enhanced, the instrument jumps to line one
of the table, then continues to process the sequence table lines.
. By default, the last line of a sequence table always jumps back to line
one, unless you have set another jump destination.
Do the following steps to set the Goto One value:
1. Move the cursor to the line in which to set the Goto One value.
2. Move the cursor to the Goto One column.
3. Push Data Entry (bottom).
4. Push Goto One (side) to toggle between On and Off.
The Off state is a blank in the column.
5. Push the CLEAR MENU on the front panel to exit the setting mode.
The Logic Jump column specifies a conditional jump to a line in the sequence
table. Conditional jumps move to a sequence line depending on the value of the
TTL logic signals on the EVENT IN rear panel connector. The instrument uses
event signals to trigger line jumps in the sequence table. Logic Jump functional-
ity is only valid when the Run Mode is set to Enhanced. Note that this setting is
neglected in the subsequence.
Goto OneLogic Jump