The Setup Menu Screen
AWG610 Arbitrary Waveform Generator User Manual 3-33
. You can change the analog output amplitude and offset values directly in
any screen by using the Vertical LEVEL/SCALE and OFFSET knobs on the
front-panel, respectively.
You can display the Setup Vertical menu at any time by pushing the VERTICAL
MENU front-panel button.
This button lets you set the output waveform band limit. You can select 20 MHz,
50 MHz, 100 MHz, 200 MHz or Through (no limiting).
Do the following steps to set the output waveform band limit:
1. Push SETUP (front)Vertical (bottom)Filter (side). The instrument
highlights the Through screen icon.
2. Use the general purpose knob to select 20 MHz, 50 MHz, 100 MHz,
200 MHz or Through.
This button lets you set the analog waveform signal output voltage range from
20 mVp-p to 2.0 Vp-p, in 1 mV increments, terminated into 50 W. You can only
get the maximum output of 2.0 Vp-p if the waveform file is using the full 8-bit
DAC range of ±1.000.
To set the marker output levels, refer to the Marker... menu description on
page 3–34.
Do the following steps to set the waveform output levels:
1. Push SETUP (front)Vertical (bottom)Amplitude (side). The instrument
highlights the Amplitude screen icon.
2. Use the general purpose knob, numeric buttons, keyboard, or LEVEL/
SCALE knob to set the output amplitude value. If you use a knob, use the
a or ' button to select the digit to change.
This button lets you set the waveform output offset voltage. You may set any
value from –1.000 to 1.000 V in 1 mV increments. The VERTICAL:OFFSET
knob on the front panel works in every display except Quick Edit.
Do the following steps to set the waveform offset value:
1. Push SETUP (front)Vertical (bottom)Offset (side). The instrument
highlights the Offset screen icon.
2. Use the general purpose knob, numeric buttons, keyboard, or VER-
TICAL:OFFSET knob to change the offset value. If you use the general
purpose knob, use the a or ' buttons or keys to select the digit to change.
FilterAmplitudeOffset