Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual B-29
Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out)These procedures check the accuracy of the AWG610 Arbitrary Waveform
Generator direct waveform outputs; amplitude and offset.
Equipment
required
A 50ĂWBNCĂcoaxial cable, a 50ĂWĂprecision terminator,an
SMA(Fe)-BNC(Ma) adapter, BNC(female)ĆtoĆdual banana
adapter,and a digital multimeter (DMM).
Prerequisites The AGW610 Arbitrary WaveformGenerator must meet the
prerequisites listedon page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W BNC coaxial cable, an SMA(Fe)-BNC(Ma) adapter, a 50 W
precision terminator, and a BNC-to-dual banana adapter to connect the
AWG610 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B–10).
DMM
50 WBNCĂcoaxial cable
InputĂconnector
+ĂBNCĆtoĆdual banana adapter
+Ă50 WBNCĂcoaxialĂcable
AWG610Arbitrary Waveform Generator
CH1ĂoutputĂconnector
+ĂSMA(Fe)ĆBNC(Ma) adapter
+ĂBNC 50 WĂprecision terminator
+Ă50 WĂBNC coaxial cable
Figure B-10: Direct DA output amplitude accuracyinitial test hookup
2. Set the DMM controls as follows:
Mode .......................... VDC
Range ......................... 2V
Input . ......................... Front