Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual B-15AWG610Test Record (cont.)
Instrument Serial Number:ą Certificate Number:ą
Temperature:ą RH%:ą
Date of Calibration:ą Technician:ą
AWG610Performance Test MaximumOutgoingIncomingMinimum
TriggerInput
Positive Threshold
(set the triggering level to 5 V)
No trigger when input voltage vary
from0Vto4.65 V.
Triggerwhen input voltage vary
from 4.65 V to 5.35 V.
Pass/Fail
Pass/Fail
__________
__________
__________
__________
Pass/Fail
Pass/Fail
Negative Threshold
(set the triggering level to -5 V)
No trigger when input voltage vary
from0Vto-4.65V.
Triggerwhen input voltage vary
from -5.35 V to -4.65 V.
Pass/Fail
Pass/Fail
__________
__________
__________
__________
Pass/Fail
Pass/Fail
Event Input and Enhanced Mode
Check Event Input with Strobe Off
pin 0 (when SW1 of the ground closure is closed )
pin 1 (when SW2 of the ground closure is closed)
pin 2 (when SW3 of the ground closure is closed)
pin 3 (when SW4 of the ground closure is closed)
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
__________
__________
__________
__________
__________
__________
__________
__________
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Check Strobe Input
(when SW5 of the ground closure is on and off) Pass/Fail __________ __________ Pass/Fail
1/4 Clock Frequency and 10 MHz Reference Input
Check 1/4 Clock Out Frequency
Internal Clock = 200 mH 49999 950 Hz __________ __________ 50 000 050 Hz
Check output with 10 MHz Reference Input
10 mH Ref In = 10 mH 49 990 000 Hz __________ __________ 50 010 000 Hz
Check output with 10.1 MHz Reference Input
10 mH Ref In = 10.1 mH 50490 000 Hz __________ __________ 50 510 000 Hz
1/4 Clock Output Level
Check 1/4 Clock Amplitude (Larger than 0.50 V) Pass/Fail __________ __________ Pass/Fail
10 MHz Reference Output Level
Check 1/4 Clock Amplitude ≥1.00 V __________ __________ N/A