Automatic control couples the attenuation to the reference level. Note that setting the attenuation using this command automatically sets the attenuation coupling to “MANUAL”, thereby allowing it to be defined independently of the reference level.

Bytes to Follow: 1 byte

1)Attenuation (0 – 51)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid attenuation

238 (EEh) Time Out Error

Set AM/FM Demodulation – Control Byte #145 (91h)

Description: Sets the AM/FM/SSB Demodulation state. This command is also used to set the type of Modulation, volume, Demodulation Frequency, BFO Adjust (SSB only) and the Demodulation time. On turning demodulation ON, after exiting remote, at the end of every sweep, demodulation is performed at the Demodulation frequency for a period of time specified in the Demod Time.

Bytes to Follow: 16 bytes

1)Set AM/FM/SSB Demod On/Off275

2)Demodulation Type276

3)Speaker Volume (higher byte)277

4)Speaker Volume (lower byte)

5)Demodulation Time278 (highest byte)

6)Demodulation Time

7)Demodulation Time

8)Demodulation Time (lowest byte)

9)Demodulation Frequency279 (highest byte)

10)Demodulation Frequency

11)Demodulation Frequency

12)Demodulation Frequency (lowest byte)

13)SSB BFO Adjust280 (highest byte)

14)SSB BFO Adjust

15)SSB BFO Adjust

16)SSB BFO Adjust (lowest byte)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error

238 (EEh) Time Out Error

27500h = Off, 01h = On

27600h = FM Wideband, 01h = FM Narrowband, 02h = AM, 03h = SSB Lower, 04h = SSB Upper

277Speaker Volume is from 0 to 100 in steps of 10

278Demodulation time in milliseconds from 100 millisec to 500 seconds

279If Option 6 is installed and the frequency converter module is attached, the frequencies should be scaled by the scale factor of the module. If the module is not attached, the frequencies are sent in Hz. Use Control Word A203 to determine whether a module is attached and the appropriate scale factor.

280BFO Valid Values are -10 kHz to +10 kHz. Send value as BFO (in Hz) + 10,000. For Example -10 kHz would be sent as 0, 0 would be sent as 10000 and +10 kHz would be 20000