Read InstaCal Module ASCII Serial Number – Control Byte #124 (7Ch)

Description: Returns the InstaCal Module serial number in ASCII.

Bytes to Follow: 1 byte

1)Serial number storage location (01h=main serial, 02h=secondary) Site Master Returns: 8 bytes

1-8) Serial Number, in ASCII

Set Site Master Marker (Peak/Valley) – Control Byte #129 (81h)

Description: Sets an individual marker in current measurement mode to either peak (maximum) signal or valley (minimum) signal.

Bytes to Follow: 2 bytes

1)Marker Number (01h = marker 1, 02h = marker 2, 03h = marker 3, 04h = marker 4, 05h = marker 5, 06h = marker 6)

2)Marker Line Search Status (01h = Peak, 00h = Valley )

Site Master Returns: 3 bytes (1 byte if an error occurs)

1)Marker Position (higher byte)274

2)Marker Position (lower byte)

3)255 (FFh) Operation Complete Byte

or

1)224 (E0h) Parameter Error : Invalid marker or marker search status 238 (EEh) Time Out Error

Set / Reset Spectrum Analyzer External Reference – Control Byte #133 (85h)

Description: Sets the external reference frequency for the spectrum analyzer in increments of 1 MHz from 2 – 20 MHz. The frequencies are sent in Hz.

Bytes to Follow: 1 byte if turning the reference OFF, 5 bytes if turning the reference ON

Turn OFF the external reference:

1)00h - Turn OFF the frequency reference

or

Turn ON the external reference (the reference frequency is also sent):

1)01h - Turn ON the frequency reference

2)External Reference Frequency (in Hz) (highest byte)

3)External Reference Frequency (in Hz)

4)External Reference Frequency (in Hz)

5)External Reference Frequency (in Hz) (lowest byte)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error

238 (EEh) Time Out Error

274The marker position is sent as a data point on the display. Equivalent Frequency = (position * span / (# data points - 1)) + start frequency

Site Master PM

91