89)Single Limit149 (highest byte)

90)Single Limit

91)Single Limit

92)Single Limit (lowest byte)

93)Multiple Limit Segment # (1)

94)Multiple Limit Segment Status

95)Multiple Limit Start X150 (highest byte)

96)Multiple Limit Start X

97)Multiple Limit Start X

98)Multiple Limit Start X (lowest byte)

99)Multiple Limit Start Y (higher byte)

100)Multiple Limit Start Y (lower byte)

101)Multiple Limit End X (highest byte)

102)Multiple Limit End X

103)Multiple Limit End X

104)Multiple Limit End X (lowest byte)

105)Multiple Limit End Y (higher byte)

106)Multiple Limit End Y (lower byte) 107–162) Repeat bytes 93-106 for segments 2-5

163)Start Distance151 (highest byte)

164)Start Distance

165)Start Distance

166)Start Distance (lowest byte)

167)Stop Distance (highest byte)

168)Stop Distance

169)Stop Distance

170)Stop Distance (lowest byte)

171)Distance Marker 1152 (higher byte)

172)Distance Marker 1 (lower byte)

173)Distance Marker 2 (higher byte)

174)Distance Marker 2 (lower byte)

175)Distance Marker 3 (higher byte)

176)Distance Marker 3 (lower byte)

177)Distance Marker 4 (higher byte)

178)Distance Marker 4 (lower byte)

179)Distance Marker 5 (higher byte)

180)Distance Marker 5 (lower byte)

181)Distance Marker 6 (higher byte)

182)Distance Marker 6 (lower byte)

183)Relative Propagation Velocity153 (highest byte)

184)Relative Propagation Velocity

185)Relative Propagation Velocity

186)Relative Propagation Velocity (lowest byte)

187)Cable Loss154 (highest byte)

188)Cable Loss

149See Control Byte #6 “Set Site Master VNA Single Limit” for data format.

150See Control Byte #112 “Set Site Master VNA Segmented Limit Lines” for data format.

151Distance data uses units 1/100,000m (or feet)

152Marker Point = ( # data points – 1 ) * ( marker dist – start dist ) / ( stop dist – start dist )

153Relative Propagation Velocity uses units 1/100,000

154Cable Loss uses units 1/100,000 dB/m or 1/100,000 dB/ft.