Write Antenna – Control Byte #82 (52h)
Description: Receives an antenna to the Site Master via the serial port.
An antenna is described with an index into the list
Frequencies are sent in Hz, scaled by the scale factor.
The value of the antenna factor should be sent as (value * 100).
Bytes to Follow: 26 – 380, depending on the number of antenna factors
1)Antenna List Index
18)Number of Antenna Factors (max = 60)
For each antenna factor:
1)Frequency (highest byte)
2)Frequency
3)Frequency
4)Frequency (lowest byte)
5)Antenna Factor (higher byte)
6)Antenna Factor (lower byte)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error
238 (EEh) Time Out Error
Recall Antenna – Control Byte #83 (53h)
Description: Sends an antenna from the Site Master via the serial port.
An antenna is described with an index into the list
Frequencies are sent in Hz, scaled by the scale factor.
The value of the antenna factor should be sent as (value * 100).
Bytes to Follow: 1 byte
1) Antenna List index
Site Master Returns:
1)Maximum Antenna Number (10)
18)Number of Antenna Factors (max = 60)
1)Frequency (highest byte)
2)Frequency
3)Frequency
4)Frequency (lowest byte)
78 | Site Master PM |