5)Antenna Factor (higher byte)

6)Antenna Factor (lower byte)

Set Field Strength Measurement – Control Byte #84 (54h)

Description: Sets the state of the measurement (ON or OFF) and the antenna index for the field strength measurement. Antennas 1-10 are custom antennas. Antennas 11-15 are standard antennas. The standard antennas are:

11.Anritsu # 2000-1030 (MAXRAD MPA1750 - 1710-1880 MHz)

12.Anritsu # 2000-1031 (MAXRAD MPA1850 - 1850-1990 MHz)

13.Anritsu # 2000-1032 (MAXRAD MPA2450 - 2400-2483.5 MHz)

14.Anritsu # 2000-1200 (Centurion EXC SM806 - 806-899 MHz)

15.Anritsu # 2000-1035 (Centurion EXE-902-SM - 896-941 MHz)

If the FCN4760 frequency converter module is attached, the standard antenna is:

11. Anritsu # 2000-1361 (5725-5825 MHz)

Note that if the field strength measurement is turned ON, all other measurements (channel power, adjacent channel power) are turned OFF.

Bytes to Follow: 2 bytes

1)Field Strength Measurement State (On/Off)

2)Antenna List index (1-15)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid state or index

238 (EEh) Time Out Error

Set Channel Power – Control Byte #85 (55h)

Description: Sets the state of the measurement (ON or OFF), and the setup parameters to perform the channel power measurement.

Send a 0 (zero) following the command to set the channel power measurement in the current setup.

Send a 1 (one) to set the channel power associated with the trace that was most recently uploaded by command #36, Upload Sweep Trace.

If Option 6 is installed and the frequency converter module is attached, the frequencies should be scaled by the scale factor of the module. If the module is not attached, frequencies are sent in Hz. Use Control Word A203 to determine whether a module is attached and the appropriate scale factor.

Note that if the channel power measurement is turned ON, all other measurements (field strength, adjacent channel power) are turned OFF.

Bytes to Follow: 14 bytes

1)Channel Power Location (0 = current setup, 1 = last uploaded trace)

2)Channel Power Measurement State (On/Off)

3-6) Center Frequency

7-10) Integration Bandwidth

11-14) Span Frequency

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error

238 (EEh) Time Out Error

Site Master PM

79

Page 85