1)Fail Counter (higher byte)

2)Fail Counter (lower byte)

3)255 (FFh) Operation Complete Byte

or

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Module not attached

238 (EEh) Time Out Error

Clear Module Fail Counter – Control Word (A205h)

(Option 6 only)

Description: Sets the module lock fail counter to 0.

Bytes to Follow: 0 bytes

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Module not attached

238 (EEh) Time Out Error

Remote Self Test - Control Word (AA15h)

Description: Trigger the equivalent of a "key press" selftest.

NOTE: The response bytes will not all be returned immediately. The first 12 will be returned, then there will be a slight delay before the next 14 are returned, then a final delay while the T1/E1 selftest is per- formed and the final 12 bytes are returned.

Bytes to Follow: 0 bytes

Site Master Returns:

S331D (No Options/Option 3): 25 bytes

1)Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2￿ C) (higher byte)

2)Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2￿ C) (lower byte)

3)Memory Check (01h: Pass, 00h: Fail)

4)RTC Voltage Check (01h: Pass, 00h: Fail)

5)Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (higher byte)

6)Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (lower byte)

7)VNA Lock Failure Counter (higher byte)

8)VNA Lock Failure Counter (lower byte)

9)VNA Integrator Failure Counter (higher byte)

10)VNA Integrator Failure Counter (lower byte)

11)SPA LO Failure Counter (higher byte)

12)SPA LO Failure Counter (lower byte)

13)H/W Config - Mother Board ID

14)H/W Config - SPA Board ID

15)H/W Config - T1E1 Board ID

16)H/W Config - PLD1 ID

17)H/W Config - PLD2 ID

18)H/W Config - T1E1 COLD ID

19)VNA PLL Lock Failure Test - Status (01h: Pass, 00h: Fail)

20)VNA PLL Lock Failure Test - Failed data point # (Ignore this byte if the Lock Fail Test Status was Pass)

21)VNA PLL Lock Failure Test - Failed PLL # (Ignore this byte if the Lock Fail Test Status was Pass)

22)VNA Integration Test - Status (01h: Pass, 00h: Fail)

23)VNA Integration Test - Failed data point # (Ignore this byte if the Integration Test Status was Pass)

Site Master PM

105

Page 111
Image 111
Anritsu S332D, S331D manual Clear Module Fail Counter Control Word A205h Option 6 only, Remote Self Test Control Word AA15h