1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error
238 (EEh)
Set Spectrum Analyzer Resolution Bandwidth – Control Byte #141 (8Dh)
NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #106 to access the RBW.
Description: Sets the resolution BW frequency for the Spectrum Analyzer.
Bytes to Follow: 4 bytes
1)Resolution Bandwidth (frequency in Hz) (highest byte)
2)Resolution Bandwidth (frequency in Hz)
3)Resolution Bandwidth (frequency in Hz)
4)Resolution Bandwidth (frequency in Hz) (lowest byte)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid RBW
238 (EEh) Time Out Error
Set Spectrum Analyzer Video Bandwidth – Control Byte #142 (8Eh)
NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #107 to access the VBW.
Description: Sets the video BW frequency for the Spectrum Analyzer.
Bytes to Follow: 4 bytes
1)Video Bandwidth (frequency in Hz) (highest byte)
2)Video Bandwidth (frequency in Hz)
3)Video Bandwidth (frequency in Hz)
4)Video Bandwidth (frequency in Hz) (lowest byte)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid VBW
238 (EEh) Time Out Error
Set Spectrum Analyzer Attenuation – Control Byte #143 (8Fh)
NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #111 to access the Attenuation.
Description: Sets the attenuation of the Spectrum Analyzer. Send a value of 255 (FFh) to enable dynamic attenuation.
Site Master PM | 93 |