Anritsu S332D, S331D manual Set Spectrum Analyzer Video Bandwidth Control Byte #142 8Eh

Models: S332D S331D

1 132
Download 132 pages 8.91 Kb
Page 99
Image 99

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error

238 (EEh) Time-out Error

Set Spectrum Analyzer Resolution Bandwidth – Control Byte #141 (8Dh)

NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #106 to access the RBW.

Description: Sets the resolution BW frequency for the Spectrum Analyzer.

Bytes to Follow: 4 bytes

1)Resolution Bandwidth (frequency in Hz) (highest byte)

2)Resolution Bandwidth (frequency in Hz)

3)Resolution Bandwidth (frequency in Hz)

4)Resolution Bandwidth (frequency in Hz) (lowest byte)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid RBW

238 (EEh) Time Out Error

Set Spectrum Analyzer Video Bandwidth – Control Byte #142 (8Eh)

NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #107 to access the VBW.

Description: Sets the video BW frequency for the Spectrum Analyzer.

Bytes to Follow: 4 bytes

1)Video Bandwidth (frequency in Hz) (highest byte)

2)Video Bandwidth (frequency in Hz)

3)Video Bandwidth (frequency in Hz)

4)Video Bandwidth (frequency in Hz) (lowest byte)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid VBW

238 (EEh) Time Out Error

Set Spectrum Analyzer Attenuation – Control Byte #143 (8Fh)

NOTE: This command is new to the Site Master S33xD models. Use it instead of Control Byte #111 to access the Attenuation.

Description: Sets the attenuation of the Spectrum Analyzer. Send a value of 255 (FFh) to enable dynamic attenuation.

Site Master PM

93

Page 99
Image 99