Power Meter Relative Mode On/Off – Control Byte #41 (29h) | (Option 29 only) |
Description: Enable or disable Power Meter Relative Mode.
Bytes to Follow: 1 byte
1)Relative Mode State
00h = Off
01h = On w/ trigger (use the current power level as a reference power level)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid parameter
238 (EEh)
Power Meter Offset Mode On/Off – Control Byte #42 (2Ah) | (Option 29 only) |
Description: Enable or disable Power Meter Offset Mode.
Bytes to Follow: 5 bytes
1) On/Off (01h = On, 00h = Off)
2 - 5) Offset Power level in dB (Multiplied by 1000)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid parameter
238 (EEh)
Note:If you turn the Offset mode off, you must still send the other bytes. Bytes 2 - 5 will be ignored.
Power Meter Zero Mode On/Off – Control Byte #43 (2Bh) | (Option 29 only) |
Description: Enable or disable Power Meter Zeroing Mode.
Bytes to Follow: 1 byte
1)Zero Mode Status 00h = Off
01h = On with trigger (current power level is referenced as
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid status
238 (EEh)
Power Meter RMS Averaging On/Off – Control Byte #44 (2Ch) | (Option 29 only) |
Description: Set Power Meter RMS Averaging. Enabling to one of three different levels, or off.
Bytes to Follow: 1 byte
1)RMS Averaging State 00h = Off
01h = On (Low) with trigger (current power level is referenced as
03h = On (High)
Site Master Returns: 1 byte
Site Master PM | 73 |