1)Status Byte

255 (FFh) Operation Complete Byte

Select SPA/Power Meter Signal Standard - Control Word (A103h)

Description: Selects a Signal Standard. Use this command for both Spectrum Analyzer and Power Meter modes. See the “Signal Standards” section for a list of standards and their indices.

Bytes to Follow: 1 byte

1)Signal Standard - See the “Signal Standards” section for a list of standards and their indices.

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid signal standard

238 (EEh) Time Out Error

Select SPA/Power Meter Channel - Control Word (A104h)

Description: Selects a channel within the range of the currently selected signal standard. Use this command for both Spectrum Analyzer and Power Meter modes. See the “Signal Standards” section for a list of valid channels for the selected channel.

Bytes to Follow: 2 bytes

1)Channel (higher byte)

2)Channel (lower byte)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid channel

238 (EEh) Time Out Error

Read External Module Name – Control Word (A201h)

(Option 6 only)

Description: Returns the name of the attached external converter module (Option 6).

For example, module name “FCN4760” will be received as: c,46,43,4e,34,37,36,30,0,0,0,0,0,ff

Bytes to Follow: 0 bytes

Site Master Returns: 14 bytes (success) OR 1 byte (failure)

1)Length of Name (12) 2-13) Module Name

14)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Module not attached

238 (EEh) Time Out Error

or

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Module not attached

238 (EEh) Time Out Error