Site Master Returns: 1 byte
1)Status Byte
255 (FFh) Operation Complete Byte
Select SPA/Power Meter Signal Standard - Control Word (A103h)
Description: Selects a Signal Standard. Use this command for both Spectrum Analyzer and Power Meter modes. See the “Signal Standards” section for a list of standards and their indices.
Bytes to Follow: 1 byte
1)Signal Standard - See the “Signal Standards” section for a list of standards and their indices.
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid signal standard
238 (EEh) Time Out Error
Select SPA/Power Meter Channel - Control Word (A104h)
Description: Selects a channel within the range of the currently selected signal standard. Use this command for both Spectrum Analyzer and Power Meter modes. See the “Signal Standards” section for a list of valid channels for the selected channel.
Bytes to Follow: 2 bytes
1)Channel (higher byte)
2)Channel (lower byte)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid channel
238 (EEh) Time Out Error
Read External Module Name – Control Word (A201h) | (Option 6 only) |
Description: Returns the name of the attached external converter module (Option 6).
For example, module name “FCN4760” will be received as: c,46,43,4e,34,37,36,30,0,0,0,0,0,ff
Bytes to Follow: 0 bytes
Site Master Returns: 14 bytes (success) OR 1 byte (failure)
1)Length of Name (12)
14)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Module not attached
238 (EEh) Time Out Error
or
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Module not attached
238 (EEh) Time Out Error
Site Master PM | 103 |