Option 6: | “6/” |
Option 10: | “10/” |
Option 21: | “21/” |
Option 29: | “29/” |
Option 50: | “50/” |
If no options are installed: “None”
Query Power Level – Control Byte #39 (27h) | (Option 29 only) |
Description: Return Power Level at the RF In port. Also returns power meter settings. |
|
Bytes to Follow: 0 bytes |
|
Site Master Returns: 30 bytes |
|
1) Status Byte # 1(0b = Off, 1b = On)
(LSB) bit 0 : Unit (0b - Watt/%, 1b - dBm/dB)
bit 2 : Relative Mode On/Off
bit 3: Offset Mode On/Off
bit 4: Zero Mode On/Off
bits
2)RMS Averaging Status263
3 | - 6) | Relative Mode Reference Power Level in dBm |
7 | - 10) | Offset Mode Power Level |
11 - 14) | Zero Mode Power Level | |
15 - 18) | Absolute Power Level | |
19 - 22) | Power | |
23 - 26) | Center Frequency | |
27 - 30) | Span Frequency |
Notes:
Power is returned as (dBm * 1000)
Relative power is returned as (dB * 1000)
Offset is returned as (dB * 1000)
Frequencies are scaled by the scale factor
Set Power Meter Units – Control Byte #40 (28h) | (Option 29 only) |
Description: Set Power Meter units to Watts or dBm.
Bytes to Follow: 1 byte
1)Units
00h = Watt (% if in relative mode)
01h = dBm (dB if in relative mode)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Invalid Units
238 (EEh)
263 RMS Averaging: 00h = Off, 01h = Low, 02h = Medium, 03h = High
72 | Site Master PM |