Anritsu S332D, S331D manual Set Spectrum Analyzer Peak Hold Control Byte #105 69h

Models: S332D S331D

1 132
Download 132 pages 8.91 Kb
Page 91
Image 91

Set Spectrum Analyzer Peak Hold – Control Byte #105 (69h)

Description: Sets the max hold and min hold settings on the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)Peak Hold State

00h – Peak Hold Off

01h – Max Hold On

02h – Min Hold On

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid state

238 (EEh) Time Out Error

Set Spectrum Analyzer Resolution Bandwidth – Control Byte #106 (6Ah)

NOTE: This command exists for backward compatibility with the S33xC models. Features new to the S33xD models are not available here. To access the new features use Control Byte #141 (8Dh).

Description: Sets the resolution BW frequency for the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)Resolution Bandwidth Index

00h – 10 kHz BW

01h – 30 kHz BW

02h – 100 kHz BW

03h – 1 MHz BW

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid RBW Index

238 (EEh) Time Out Error

Set Spectrum Analyzer Video Bandwidth – Control Byte #107 (6Bh)

NOTE: This command exists for backward compatibility with the S33xC models. Features new to the S33xD models are not available here. To access the new features use Control Byte #142 (8Eh).

Description: Sets the video BW frequency for the Spectrum Analyzer.

Bytes to Follow: 1 byte

1)Video Bandwidth Index 00h – 100 Hz BW 01h – 300 Hz BW 02h – 1 kHz BW 03h – 3 kHz BW 04h – 10 kHz BW 05h – 30 kHz BW 06h – 100 kHz BW 07h – 300 kHz BW

Site Master Returns: 1 byte

Site Master PM

85

Page 91
Image 91
Anritsu S332D, S331D manual Set Spectrum Analyzer Peak Hold Control Byte #105 69h