Notes:

Battery Voltage in 1/10th of a Volt (e.g. 124 = 12.4 Volts)

Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2￿C) or degree Fahrenheit (e.g., 934 = 93.4￿F), depending on the current measurement unit (Metric or English) selected

Read Fail Counter – Control Byte #22 (16h)

Description: Reads the Fail Counter. Values are integer numbers of failures.

Bytes to Follow: 0 bytes

Site Master Returns: 8 bytes

1)Value of SM Lock Fail Counter (higher byte)

2)Value of SM Lock Fail Counter (lower byte)

3)Value of Integration Fail Counter (higher byte)

4)Value of Integration Fail Counter (lower byte)

5)Value of SA Lock Fail Counter (higher byte)

6)Value of SA Lock Fail Counter (lower byte)

7)Value of SA Fatal Error Counter (higher byte)

8)Value of SA Fatal Error Counter (lower byte)

Query Trace Names – Control Byte #24 (18h)

Description: Returns a list of all saved traces.

Bytes to Follow: 0 bytes

Site Master Returns: 3 + (41 x number of saved traces) bytes

1-2) # of saved traces For each trace:

1-2) Trace Index

3)Measurement Mode (refer to Control Byte #3)

4-21) Date/Time in string format (“MM/DD/YYYYHH:MM:SS”)

22-25) Date/Time as Unsigned Long Integer (Seconds Since January 1, 1970)

26-41) Trace Name (16 bytes)

255 (FFh) Operation Complete Byte

Delete Sweep Trace – Control Byte #25 (19h)

Description: Delete single trace or all stored sweep traces in Site Master.

Bytes to Follow: 1 byte

1)0 - Delete all traces

X - Delete single trace #X

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

Site Master PM

29

Page 35
Image 35
Anritsu S332D Read Fail Counter Control Byte #22 16h, Query Trace Names Control Byte #24 18h, Bytes to Follow 0 bytes