Store Sweep Trace – Control Byte #16 (10h)

Description: Saves current trace to the next available memory location. Trace name can be set using control byte #9, “Set Reference Number” before executing this command.

Bytes to Follow: 0 bytes

Site Master Returns: 5 bytes

1-4) Time/Date Stamp (In long integer format)

5)Operation result:

255 (FFh) Operation Complete Byte

224 (E0h) Out of memory (Memory full)

238 (EEh) Time-out Error

Recall Sweep Trace – Control Byte #17 (11h)

NOTE: This command exists for backward compatibility with the S33xC models. Features new to the S33xD models are not available here. To access the new features, use Control Byte #33 (21h). This command cannot be used with a frequency converter module attached (Option 6 required).

Description: Queries the Site Master for sweep trace data.

Note: Before you can recall a sweep stored in non-volatile memory (trace numbers 1-200) you must build a trace table in the Site Master’s RAM. Use Control Byte #24 to build the trace table. Since the trace table exists in RAM, Control Byte #24 must be executed every time the Site Master’s power is cycled.

Bytes to Follow: 1 byte

0 = Last sweep trace before entering remote mode (sweep trace in RAM) 1- 200 = Specific saved sweep number (stored sweeps in Flash memory)

Site Master Returns:

1-2)

# of following bytes (total length - 2)

3-4)

 

Not Used

5-11)

 

Model Number (7 bytes in ASCII)

12-15)

Software Version (4 bytes ASCII)

16)

Measurement Mode3

17-20)

Time/Date (in Long Integer4)

21-30)

Date in String Format (mm/dd/yyyy)

31-38)

Time in String Format (hh:mm:ss)

39-54)

Reference number stamp (16 bytes in ASCII)

55-56)

# data points (130, 259, 517 or 400)

For all Site Master Modes:

57)Start Frequency5 (highest byte)

58)Start Frequency

59)Start Frequency

60)Start Frequency (lowest byte)

61)Stop Frequency (highest byte)

62)Stop Frequency

3Refer to Control Byte #3 “Select Measurement Mode” for detailed value.

4Time/Date long integer representation is in seconds since January 1, 1970

5Frequency units are Hz

14

Site Master PM

Page 20
Image 20
Anritsu S331D, S332D Store Sweep Trace Control Byte #16 10h, Recall Sweep Trace Control Byte #17 11h, Site Master Returns