Anritsu S331D, S332D manual Save System Setup Control Byte #18 12h

Models: S332D S331D

1 132
Download 132 pages 8.91 Kb
Page 26
Image 26

300)Status Byte 5

( 0b = Off/Beep if data is below line, 1b = On/Beep if data is above line)

(LSB) bit 0 : Multiple Limit Upper Segment 3

Status On/Off

bit 1 : Multiple Limit Upper Segment 3 Beep Level ABOVE/BELOW

bit 2 : Multiple Limit Upper Segment 4

Status On/Off

bit 3 : Multiple Limit Upper Segment 4 Beep Level ABOVE/BELOW

bit 4

: Multiple Limit Upper Segment 5

Status On/Off

bit 5

: Multiple Limit Upper Segment 5 Beep Level ABOVE/BELOW

bit 6

: Multiple Limit Lower Segment 1 Status On/Off

bit 7

: Multiple Limit Lower Segment 1 Beep Level ABOVE/BELOW28

303) Status Byte 6

(0b = Off/Beep if data is BELOW line, 1b = On/Beep if data is ABOVE line)

(LSB) bit 0 : Multiple Limit Lower Segment 2

Status On/Off

bit 1 : Multiple Limit Lower Segment 2 Beep Level ABOVE/BELOW

bit 2 : Multiple Limit Lower Segment 3

Status On/Off

bit 3 : Multiple Limit Lower Segment 3 Beep Level ABOVE/BELOW

bit 4

: Multiple Limit Lower Segment 4

Status On/Off

bit 5

: Multiple Limit Lower Segment 4 Beep Level ABOVE/BELOW

bit 6

: Multiple Limit Lower Segment 5

Status On/Off

bit 7

: Multiple Limit Lower Segment 5 Beep Level ABOVE/BELOW

304)Status Byte 7

bits 0-6: Number of sweeps to average (1-25, 1 implies no averaging)

bit 7: Not Used

305)Reference Level Offset 29(highest byte)

306)Reference Level Offset

307)Reference Level Offset

308)Reference Level Offset (lowest byte) 309-338) Not Used

339-1938) Sweep Data

(400 points * 4 bytes/point= 1600 bytes)

4 bytes for each data point

1.dBm30 MSB

2.dBm

3.dBm

4.dBm LSB

Site Master Returns (For invalid sweeps/empty stored sweep locations): 11 bytes 1-2) Number of following bytes (9 bytes for invalid sweep recall)

3-4) Model # (unsigned integer, 14h for Site Master S33xD)

5-11) Extended Model # (7 bytes in ASCII)

Site Master Returns (Invalid sweep location): 1 byte

1) 224 (E0) Parameter Error: Invalid sweep location

Save System Setup – Control Byte #18 (12h)

Description: Saves current system setup parameters to a specific setup store location.

The Site Master saves all parameters described in Query System Status - control byte #29 (1Dh), (except Serial Port Echo Status) to the specified store location. Store location 0 is the run-time setup of the Site Master. It holds the power-on defaults of the Site Master.

28LOWER limits always trigger an error beep if data is BELOW the limit segment, for example, this bit is always 0b.

29Value sent as ( value in dBm * 1000 ) + 270,000

30Value sent as ( value in dBm * 1000 ) + 270,000