Write Custom Cable – Control Byte #80 (50h)
Description: Write a cable parameter in the custom cable list.
Bytes to Follow: 25 bytes
1)Not Used
2)Cable List index (0 - 49)
3 – 17) Cable Description (string)
18)Propagation Velocity (highest byte)264
19)Propagation Velocity
20)Propagation Velocity
21)Propagation Velocity (lowest byte)
22)Insertion Loss (highest byte)265
23)Insertion Loss
24)Insertion Loss
25)Insertion Loss (lowest byte)
Site Master Returns: 1 byte
1)255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error
238 (EEh) Time Out Error
Recall Custom Cable – Control Byte #81 (51h)
Description: Query a cable in the custom cable list.
Bytes to Follow: 2 bytes
1)Not Used
2)Cable list index
Site Master Returns: 24 bytes
1)Upper bound of Custom Cable Index 2 – 16) Cable Description (string)
17)Propagation Velocity (highest byte)266
18)Propagation Velocity
19)Propagation Velocity
20)Propagation Velocity (lowest byte)
21)Insertion Loss (highest byte)267
22)Insertion Loss
23)Insertion Loss
24)Insertion Loss (lowest byte)
264Cable loss uses units 1/100,000 dBm or 1/100,000 dB/ft.
265Relative Propagation Velocity uses units 1/100,000.
266Relative Propagation Velocity uses units 1/100,000.
267Cable loss uses units 1/100,000 dB/m or 1/100,000 dB/ft.
Site Master PM | 77 |