1)Clock Source 00h: Internal

01h: External

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid clock source

238 (EEh) Time Out Error

Set T1/E1 Pattern – Control Word (A003h)

(Option 50 only)

Description: Sets the data pattern of T1/E1 measurement mode.

Bytes to Follow: 2 bytes

1)Data Pattern

00h: AUTO_DETECT

01h: PRBS_9

02h: PRBS_11

03h: PRBS_15

04h: PRBS_20 (O.151)

05h: PRBS_20 (O.153)

06h: PRBS_23

07h: QRSS

08h: ONE_IN_8

09h: TWO_IN_8

0Ah THREE_IN_24

0Bh: ALL_ONES

0Ch: ALL_ZEROS

0Dh: T1_DALY

0Eh: BLUE_ALARM

0Fh: YELLOW_ALARM

10h: USER_DEFINED

2)Inverted Pattern Option (01h: Inverted; 00h: Non-inverted)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid pattern index

238 (EEh) Time Out Error

Set T1/E1 Error Insert Type/Value – Control Word (A004h)

(Option 50 only)

Description: Sets the Insertion Error type and the number of errors.

Bytes to Follow: 5 bytes

1)Error Type 00h: Bit

01h: Bert

02h: BPV

03h: Framing

2)Number of Errors (highest byte)

3)Number of Errors

4)Number of Errors

5)Number of Errors (lowest byte)

Site Master Returns: 1 byte