Anritsu S331D, S332D manual Set Spectrum Analyzer Sweep Mode Control Byte #108 6Ch

Models: S332D S331D

1 132
Download 132 pages 8.91 Kb
Page 92
Image 92

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid VBW Index

238 (EEh) Time Out Error

Set Spectrum Analyzer Sweep Mode – Control Byte #108 (6Ch)

Description: Enables or disables the Single Sweep Mode during Spectrum Analyzer mode of operation.

Single Sweep Mode activates once the Site Master exits from the remote mode.

For Single Sweep Mode during Site Master VNA modes of operation see control byte #11 (0Bh).

Bytes to Follow: 1 byte

1)Sweep Mode

00h – Single Sweep

01h – Continuous Sweep

02h – Video Trigger (Span must be 0)

03h – External Trigger (Span must be 0)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid Mode

238 (EEh) Time Out Error

Set Spectrum Analyzer Marker to Peak – Control Byte #109 (6Dh)

Description: Sets the specified marker to the peak value of the sweep.

Bytes to Follow: 1 byte

1)Marker Number (1-6)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid Marker Number

238 (EEh) Time Out Error

Set Spectrum Analyzer Marker to Center – Control Byte #110 (6Eh)

Description: Sets the center frequency equal to the frequency of the specified marker.

Bytes to Follow: 1 byte

1)Marker Number (1-4)

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error: Invalid Marker Number

238 (EEh) Time Out Error

86

Site Master PM

Page 92
Image 92
Anritsu S331D, S332D manual Set Spectrum Analyzer Sweep Mode Control Byte #108 6Ch