Anritsu S332D Sequence Site Master Calibration Control Byte #13 0Dh, Site Master Returns 2 bytes

Models: S332D S331D

1 132
Download 132 pages 8.91 Kb
Page 19
Image 19

Sequence Site Master Calibration – Control Byte #13 (0Dh)

Description: Initiates a calibration step. The Site Master must be calibrated to give accurate measurements.

The command sequence must be sent in correct order. i.e. Open -> Short -> Load. You can also abort the calibration by command – “Abort” before the command - “Load” is sent. Once command - “Load” is sent, calibration is completed, and the old calibration data is lost.

The unit under test returns #255(FFh) upon receiving the command, and return #240 (F0h) when the calibration of the connected component is completed, then waits for further commands to complete the whole calibration process.

This command is designed to be executed step by step: open, short, load. Issuing any other command during this command sequence will cause undesired results.

Bytes to Follow: 1 byte

1)Calibration Step to trigger 01h = Open

02h = Short

03h = Load

04h = Not Used

05h = Abort

Site Master Returns: 2 bytes

1)255 (FFh) Operation Complete Byte

224 (E0h) Error : Invalid Cal Operation or Cal Incomplete 238 (EEh) Time-out Error

2)240 (F0h) Calibration step is completed

Set Site Master Data Points – Control Byte #14 (0Eh)

Description: Set number of measurement data points for Site Master VNA modes.

Bytes to Follow: 1 byte

1)Number of Data Points 00h = 130 Points 01h = 259 Points 02h = 517 Points

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error : Invalid number of data points 238 (EEh) Time-out Error

Set Site Master Calibration Mode – Control Byte #15 (0Fh)

Description: Set the Site Master calibration mode to OSL Cal (standard) or FlexCal.

Bytes to Follow: 1 byte

1) Calibration Mode

00h = OSL Calibration (standard)

01h = FlexCal Calibration

Site Master Returns: 1 byte

1)255 (FFh) Operation Complete Byte

224 (E0h) Parameter Error : Invalid calibration mode

238 (EEh) Time-out Error