150)Start Distance (highest byte)76

151)Start Distance

152)Start Distance

153)Start Distance (lowest byte)

154)Stop Distance (highest byte)

155)Stop Distance

156)Stop Distance

157)Stop Distance (lowest byte)

158)Distance Marker 1 (higher byte)77

159)Distance Marker 1 (lower byte)

160)Distance Marker 2 (higher byte)

161)Distance Marker 2 (lower byte)

162)Distance Marker 3 (higher byte)

163)Distance Marker 3 (lower byte)

164)Distance Marker 4 (higher byte)

165)Distance Marker 4 (lower byte)

166)Distance Marker 5 (higher byte)

167)Distance Marker 5 (lower byte)

168)Distance Marker 6 (higher byte)

169)Distance Marker 6 (lower byte)

170)Relative Propagation Velocity (highest byte)78

171)Relative Propagation Velocity

172)Relative Propagation Velocity

173)Relative Propagation Velocity (lowest byte)

174)Cable Loss (highest byte)79

175)Cable Loss

176)Cable Loss

177)Cable Loss (lowest byte)

178)Status Byte 1: ( 0b = Off , 1b = On)

(LSB)

bit 0 : Marker 1

On/Off

 

bit 1 : Marker 2

On/Off

 

bit 2 : Marker 3

On/Off

 

bit 3 : Marker 4

On/Off

 

bit 4 : Marker 5

On/Off

 

bit 5 : Marker 6

On/Off

 

bits 6-7 : Not Used

179) Status Byte 2: (0b = Off, 1b = On)

(LSB)

bit 0 : Marker 2

Delta On/Off

 

bit 1

: Marker 3

Delta On/Off

 

bit 2

: Marker 4

Delta On/Off

 

bits 3-7: Not Used

180) Status Byte 3: (0b = Off , 1b = On)80

(LSB)

bit 0

: Single Limit On/Off

 

bit 1: CW On/Off

 

bits 2-3: Not Used

 

bit 4

: InstaCal On/Off

76Distance data uses units 1/100,000m or 1/100,000 ft

77Marker point = ( # of data points – 1 ) * ( marker dist – start dist ) / ( stop dist – start dist )

78Relative Propogation Velocity uses units 1/100,000

79Cable Loss uses units 1/100,000 dB/m or 1/100,000 dB/ft

80Bits (4,5) are as follows: (0,0)=Cal Off, (0,1)=OSL Cal, (1,0) = Impossible, (1,1) = InstaCal