The attenuators used are based on the expected value or range specified. Table 3-1 shows the attenuators used as a function of the expected value and range.

Table 3-1. HP E1429 Digitizer Measurement Ranges

Input Port

<expected

Range

Measurement

20 dB

20 dB

Internal Attenuators

 

value>

setting

Range

Input

Post

6 dB

14 dB

 

(MEAS/CONF)

<range>

 

Attenuator

Attenuator

 

 

 

 

 

 

 

 

 

 

1,2,3,4

±0.1

0.10235

-0.10230 to 0.10235

off

off

off

off

 

 

 

 

 

 

 

 

1,2,3,4

±0.2

0.2047

-0.2046 to 0.2047

off

off

on

off

 

 

 

 

 

 

 

 

1,2,3,4

±0.5

0.51175

-0.5115 to 0.51175

off

off

off

on

 

 

 

 

 

 

 

 

1,2,3,4

±1.0

1.0235

-1.0230 to 1.0235

off

on

off

off

 

 

 

 

 

 

 

 

3,4

±2.0

2.047

-2.0460 to 2.0470

off

on

on

off

 

 

 

 

 

 

 

 

3,4

±5.0

5.1175

-5.115 to 5.1175

off

on

off

on

 

 

 

 

 

 

 

 

3,4

±10.0

10.235

-10.230 to 10.235

on

on

off

off

 

 

 

 

 

 

 

 

3,4

±20.0

20.470

-20.460 to 20.470

on

on

on

off

 

 

 

 

 

 

 

 

3,4

±50.0

51.175

-51.150 to 51.175

on

on

off

on

 

 

 

 

 

 

 

 

3,4

±100.0

102.35

-102.30 to 102.35

on

on

on

on

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Using the Single-Ended Input 1V Range

When taking readings of low-level signals on the single-ended input, selecting the 1V range ensures the most accurate readings. This is achieved through a path directly to the A/D, which bypasses the attenuators and amplifier which can add noise and distortion to the signal (Figure 3-3).

The single-ended input 1V range can be set by specifying an

<expected value > of 1V in the MEASure or CONFigure command, or by specifying a range of 1V in the SENSe:VOLTage command:

SENS:VOLT:RANG 1

110 Understanding the HP E1429 Digitizer

Chapter 3

Page 110