TRIGger[:STARt][:IMMediate]

[:STARt][:IMMediate]

TRIGger[:STARt][:IMMediate] will cause a reading to be taken immediately when the digitizer is in the wait-for-trigger state (ARM event has occurred), regardless of the selected trigger source. The number of triggers (set by TRIGger:STARt:COUNt) will be decremented by one. The selected trigger source remains unchanged.

There is no query form of this command.

Comments Executable when initiated: Yes

Coupled Command: No

If the trigger system is in the idle or wait-for-arm state, TRIGger:STARt:IMMediate will cause error -211,"Trigger ignored" to be generated, and no action will be taken.

Related Commands: INITiate:IMMediate

*RST Condition: none

Example Forcing a measurement to occur

ARM:SOUR IMM

Arm trigger immediately after INIT

 

command received

TRIG:SOUR HOLD

Set trigger source to hold

INIT

Initiate trigger system, trigger will go to

 

hold

TRIG

Override the hold and take a reading

[:STARt]:SOURce

TRIGger[:STARt]:SOURce <source > configures the trigger system to respond to the specified source for taking readings.

The available sources are:

BUS: The Group Execute Trigger (GET) HP-IB command or the IEEE-488.2 *TRG common command.

ECLTrg0 and ECLTrg1: The VXIbus ECL trigger lines.

DECLtrg: Samples are taken at a dual rate, using the VXIbus ECLTRG0 trigger line to pace pre-arm readings, and the ECLTRG1 trigger line to pace the post-arm readings.

276 Command Reference

TRIGger Subsystem

Page 276
Image 276
HP E1429A manual TRIGgerSTARtIMMediate, STARtSOURce