CALibration[<chan>]:STORe:AUTO

:STORe:AUTO

CALibration[<chan >]:STORe:AUTO <mode > selects whether or not the

calibration constants will be automatically stored when commands like CALibration:GAIN and CALibration:ZERO complete.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

1 or 2

none

 

 

 

 

mode

boolean

ON1OFF0

none

 

 

 

 

Comments Executable when initiated: Yes

Coupled Command: No

Enabling CALibration[<chan >]:STORe:AUTO for either channel enables it for both channels.

Related commands: CALibration:GAIN, CALibration:SECure:STATe, CALibration:ZERO, CALibration:STORe

*RST Condition: CALibration:STORe:AUTO ON

Example Turn automatic storage of calibration values off

CAL:STOR:AUTO OFF

Disable automatic storage

:VALue

CALibration[<chan >]:VALue <number > specifies the voltage level supplied at the input. This voltage value is then used in subsequent CALibration:GAIN or CALibration:ZERO commands. The HP E1429 can not distinguish between a value which is exactly full scale, and one which is an overload -- both cases would generate the same measured value. Therefore, the voltage specified for CALibration:VALue is not allowed to be closer than 10 counts from full scale (approximately 99.5% of full scale). The following table shows the allowable CALibration:VALues which are closest to full scale on their respective voltage ranges.

208 Command Reference

CALibration[<chan>] Subsystem

Page 208
Image 208