INPut[<port>][:STATe]

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

port

numeric

1234

none

 

 

 

 

impedance

numeric

50751.0E61MOHM

Ohms

 

 

 

MINimumMAXimum

 

 

 

 

 

 

Input ports 1 and 2

: MINimum selects 50Ω; MAXimum selects

75Ω.

 

: MINimum selects 1 MΩ; MAXimum selects

Input ports 3 and 4

1 MΩ.

 

 

 

 

 

 

 

 

 

Comments Executable when initiated: Yes

Coupled Command: No

Changing the input impedance while the digitizer is taking readings (initiated) may cause unexpected reading values due to relay settling during the transition.

Related commands: SENSe:FUNCtion "VOLTage <port>"

*RST Condition: INPut12:IMPedance 50 OHM, INPut34:IMPedance 1 MOHM

.

Example Setting 75 Ω input impedance

INP:IMP 75 OHM

Set 75Ω impedance (defaults to input 1)

[:STATe]

INPut[<port>][:STATe] <mode> connects/disconnects the input ports to/from the measurement signal path. All ports have a special relay which is used to isolate the input from the signal path, regardless of whether the sensor channel is currently connected to that port or not.