ARM[:STARt]:COUNt

Parameters

 

Parameter

Parameter

 

Range of

Default

 

Name

Type

 

Values

Units

 

 

 

 

 

 

 

count

numeric

 

1 through 65535

none

 

 

 

 

MINimumMAXimum

 

 

 

 

 

9.9E+37INFinity

 

 

 

 

 

 

 

 

MINimum selects 1 arm.

 

 

 

When only post-arm readings are specified, MAXimum is computed as:

 

524,288 / TRIGger:COUNt, up to a maximum of 65,535 arms . When

 

pre-arm readings have been specified, MAXimum selects 128 arms.

 

9.9E+37 is equivalent to INFinity.

 

 

 

 

 

 

Comments Executable when initiated: No

 

 

Coupled Command: Yes. This command is coupled to the total reading count (TRIGger:STARt:COUNt). An error will result if TRIGger:STARt:COUNt is too large for the specified ARM:STARt:COUNt. See the following table for the relationship between ARM:STARt:COUNt and TRIG:STARt:COUNt.

If the count is set to INFinity or 9.9E+37, the ABORt command must be used to return the trigger system to the idle state.

If pre-arm readings are specified (SENSe:SWEep:OFFSet:POINts < 0), then the maximum number of arms is 128.

When pre-arm readings are specified (SENSe:SWEep:OFFSet:POINts < 0), an ARM:STARt:COUNt > 1 causes memory to be partitioned, and limits to be placed on TRIGger:STARt:COUNt. See the following table for limits.

188 Command Reference

ARM Subsystem

Page 188