Setting the Input Impedance

The impedance of the single-ended inputs can be set to 50Ω or 75Ω. At power-on or following a reset, the impedance is set to 50Ω. The impedance is changed using the following bits of the A/D shift register.

Bit

Name

 

 

Setting

Position

 

 

 

 

 

 

 

 

6

TERM75 1

0 - Ch1

50Ω input imedance selected

 

 

1

- Ch1

75Ω input impedance select

 

 

 

 

 

11

TERM75 2

0

- Ch2

50Ω input imedance selected

 

 

1

- Ch2

75Ω input impedance select

 

 

 

 

 

 

 

 

 

 

Procedure

Enabling the 10 MHz Filter

With the bit position known, set the input impedance as required using the procedure for reading and writing to the shift register.

At power-on or following a reset, the 10 MHz filter is switched out of the signal path (disabled) of the single-ended and differential inputs. The filter can be switched into the path (enabled) using the following bits of the A/D shift register.

Bit

Name

 

 

Setting

Position

 

 

 

 

 

 

 

 

9

FILTER1

0 - Ch1 10

MHz filter disabled

 

 

1

- Ch1 10

MHz filter enabled

 

 

 

 

 

14

FILTER 2

0

- Ch2 10

MHz filter disabled

 

 

1

- Ch2 10

MHz filter enabled

 

 

 

 

 

 

 

 

 

 

Procedure

Setting the

Measurement

Range

Note

With the bit position known, disable/enable the filter as required using the procedure for reading and writing to the shift register.

The digitizer measurement range is set using a series of attenuators (Figure 3-3). Table C-1 shows the attenuator settings used to select the corresponding measurement range.

There is a 3 ms relay settling time following each range change. Samples cannot be taken during the settling time.

370 Register Programming

Appendix C

Page 370
Image 370