*OPC?

*OPC?

*OPC? causes the HP E1429 to wait for all pending operations to complete. A single ASCII “ 1" is then placed in the ou tput queue.

If STATus:OPC:INITiate OFF is set, the ASCII “ 1" will be placed in the output queue when all commands received prior to the *OPC? have been executed. If ON is set, *OPC? waits for the digitizer to return to the idle state before placing the ” 1" in the output queue. No other commands will be executed until the “ 1" is placed in the output queue.

Comments Executable when initiated: Yes

Coupled command: No

Related commands: *OPC, *WAI

*RST Condition: none

*PMC

*PMC purges all macro definitions.

Comments Use the *RMC command to purge an single macro definition.

Executable when initiated: Yes

Coupled command: No

Related commands: *DMC, *RMC

*RST Condition: none

*PUD and *PUD?

*PUD <data> stores the specified data in the HP E1429’s non-volatile calibration memory. The data must be sent in IEEE-488.2 definite or indefinite block format. Calibration security must have been previously disabled.

The query form returns the current protected user data in IEEE-488.2 definite block format. The query form may be executed regardless of the state of calibration security.

Chapter 4

Command Reference 304

Page 304
Image 304
HP E1429A manual Opc?, Pmc, PUD and *PUD?