Table B-3. HP E1429A/B Error Messages (Cont’d)

Code

Message

Description

 

 

 

1019

Data loss detected during LBUS

Readings from the digitizer A/D were lost during a

 

transfer

Local bus transfer. This error usually occurs when

 

 

multiple digitizers are used and arming/triggering

 

 

signals between them cause readings to be missed.

 

 

 

1020

Indefinite block not allowed

When executing the VINStrument:TEST:DATA

 

 

command, the data must be in the IEEE-488.2

 

 

definite length block format.

 

 

 

1021

LBUS still running, ABOR or

The HP E1429B digitizer is in the interleaved transfer

 

VINS:LBUS:RES needed

mode and the LBUS chip is still active. The chip must

 

 

be reset before the next INITiate command.

 

 

 

1022

Local bus test data size not

When executing the VINStrument:TEST:DATA

 

multiple of 4 bytes

command, the amount of data sent must be a multiple

 

 

of 4 (bytes).

 

 

 

1213

Illegal when initiated

The command can not be executed while the digitizer

 

 

is INITiated.

 

 

 

2003

Memory address incorrect

Address specified by DIAGnostic:POKE or

 

 

DIAGnostic:PEEK? is not valid.

 

 

 

2004

Invalid address for 32-bit access

Attemping a 32-bit read from an odd numbered

 

 

address.

 

 

 

2007

Bus error

Error during DIAGnostic:PEEK or DIAGnostic:POKE?

 

 

 

 

 

 

340 Useful Tables

Appendix B

Page 340
Image 340