STATus:OPC:INITiate

Subsystem Syntax

Questionable Signal Status Register

Only bits 0 (VOLTage), 2 (TIME), and 8 (CALibration) are defined. All other bits are always 0.

Bit 0 - VOLTage: Set (1) if an overload voltage is detected during a measurement. Cleared (0) otherwise.

Bit 2 -TIME: Set (1) if the divide-by-n of the reference oscillator source can not generate a sample rate that is within 1% of the rate specified by TRIG:TIMer1 or TRIG:TIMer2. Cleared (0) otherwise.

Bit 8 - CALibration: Set (1) if an error has been detected in the non-volatile calibration memory. Cleared (0) otherwise.

STATus

 

:OPC

 

:INITiate <state >

 

:OPERation:QUEStionable

 

:CONDition?

[query only]

:ENABle <unmask >

 

[:EVENt]?

[query only]

:NTRansition <unmask >

 

:PTRansition <unmask >

 

:PRESet

[no query]

:OPC:INITiate

STATus:OPC:INITiate <state > controls whether the *OPC, *OPC?, and *WAI commands will complete immediately or whether they will wait for all measurements to complete. With STATe OFF set, these commands will complete immediately, which indicates that the parser is idle and ready for the next command, even though the HP E1429 may still be in the INITiated state and taking a measurement. With STATe ON set, these commands will wait for the Pending Operation Flag set true by INITiate:IMMediate, VINStrument:CONFigure:VME:MEMory:INITiate, or VINStrument:CONFigure:LBUS:MEMory:INITiate to return false, indicating that the trigger system is in the idle state and that all measurements have completed or been aborted by the ABORt or *RST commands.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

state

boolean

OFF0ON1

none

 

 

 

 

266 Command Reference

STATus Subsystem

Page 266
Image 266