Table B-3. HP E1429A/B Error Messages (Cont’d)

Code

Message

Description

 

 

 

-230

Data corrupt or stale

Attempting to FETCh? data from the digitizer

 

 

following a reset or other digitizer configuration

 

 

change.

 

 

 

-231

Data questionable

Reading accuracy is questionable. An example is

 

 

when the expected value and resolution parameters

 

 

of the CONFigure or MEASure command are

 

 

specified. If the resolution is too fine for the expected

 

 

value, this error occurs.

 

 

 

-240

Hardware error

The command could not be executed because of a

 

 

hardware failure.

 

 

 

-270

Macro error

*RMC <name>was executed and name is not defined.

 

 

 

-271

Macro syntax error

A syntax error occurred among the commands within

 

 

the macro.

 

 

 

-272

Macro execution error

Macro program data sequence could not be executed

 

 

due to a syntax error within the macro definition.

 

 

 

-273

Illegal macro label

The macro label defined in the *DMC command was

 

 

too long, the same as a common command keyword,

 

 

or contained invalid header syntax.

 

 

 

-274

Macro parameter error

The macro definition improperly used a macro

 

 

parameter placeholder.

 

 

 

-275

Macro definition too long

The commands within the macro could not be

 

 

executed because the string or block contents were

 

 

too long.

 

 

 

-276

Macro recursion error

A macro program data sequence could not be

 

 

executed because the sequence leads to the

 

 

execution of a macro being defined.

 

 

 

-277

Macro redefinition not allowed

A macro label in the *DMC command could not be

 

 

executed because the macro label was already

 

 

defined.

 

 

 

-278

Macro header not found

A legal macro label in the *GMC? query could not be

 

 

executed because the header was not previously

 

 

defined.

 

 

 

-312

PUD memory lost

The protected user data saved by the *PUD

 

 

command has been lost.

 

 

 

-313

Calibration memory lost

The nonvolatile calibration data used by the *CAL

 

 

command has been lost.

 

 

 

-330

Self-test failed

Note the information associated with the message for

 

 

a description of the failure.

 

 

 

-350

Queue overflow

The digitizer error queue is full and additional errors

 

 

have occurred.

 

 

 

 

 

 