DC Accuracy:

±0.5% of reading ±1% of peak-to-peak full-scale

Accuracy is specified for the average of 100 readings with inputs terminated in <1 kΩ and CAL:ZERO performed within 24 hours prior to reading in a stable ambient. For temperatures outside18-28 degrees C, add the following temperature coefficients for eachdegree below

18 C or above 28 C:

Range

% of peak-to-peak full scale, per degree

0.1024

0.055

0.2048

0.035

others

0.025

Differential Inputs, supplemental characteristics

Analog Bandwidth (4.6.1) (filter off)

15 MHz on 0.1023V through 1.023V ranges, plus 10.23V range 10 MHz on 2.046V and 20.46V ranges

4 MHz on 5.115V and 51.15V ranges

2 MHz on 102.3V range

Effective bits, relative to 1V single-ended range

(For 1 volt single-ended performance, see the previous A/D section.) Typical effective bits relative to 1V single-ended:

Range

500 kHz

10 MHz

0.1

-1.4

-0.9

0.2

-0.7

-0.3

0.5

-0.2

-0.1

1.0

-0.1

-0.3

324 Specifications

Appendix A

Page 324
Image 324
HP E1429A manual Differential Inputs, supplemental characteristics, Effective bits, relative to 1V single-ended range