TRIGger[:STARt]:COUNt

[:STARt]:COUNt

TRIGger[:STARt]:COUNt <number> specifies the total number of readings which will be taken during each ARM:STARt:COUNt cycle of a measurement. This command is identical to (and coupled to) SENSe:SWEep:POINts; setting either will set both to the same count.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

12

none

 

 

 

 

number

numeric

1 7 through 16,777,215

none

 

 

MINimumMAXimum

 

 

 

9.9E+37INFinity

 

 

 

 

 

MINimum selects 1 reading.

 

 

Memory, post-arm readings only: MAXimum = 524,288 / ARM:STARt:COUNt

Memory battery enabled: MAXimum = (524,288 / ARM:STARt:COUNt) - 4

A/D converter to VME (VXI data transfer) bus: MAXimum = 16,777,215

A/D converter to Local bus: MAXimum = 16,777,215

Digitizer memory, pre- and post-arm readings: see below

ARM:STARt:COUNt

Number of

Maximum Readings

 

 

Memory

(TRIGger:STARt:COUNt)

 

 

Segments

 

 

 

 

 

 

1

1

524,288

 

 

 

 

 

2

2

262,144

 

 

 

 

3

- 4

4

131,072

 

 

 

 

5

- 8

8

65,536

 

 

 

9 - 16

16

32,768

 

 

 

 

17

- 32

32

16,384

 

 

 

 

33

- 64

64

8,192

 

 

 

65 - 128

128

4096

 

 

 

 

NOTE: If the non-volatile mode of memory is enabled (MEMory:BATTery:STATe ON), then all of the maximum reading counts shown above decrease by four. These four memory locations in each segment hold the data necessary to recover all readings after a power failure.