CALibration[<chan>]:VALue

Maximum Gain

Voltage Range

Allowable

Calibration Values (V)

(Volts)

Ports

 

 

 

-.10180 and .10185

0.10235

1,2,3,4

 

 

 

-.2036 and .2037

0.2047

1,2,3,4

 

 

 

-.5090 and .50925

0.51175

1,2,3,4

 

 

 

-1.0180 and 1.0185

1.0235

1,2,3,4

 

 

 

-2.036 and 2.037

2.037

3,4

 

 

 

-5.090 and 5.0925

5.1175

3,4

 

 

 

-10.180 and 10.185

10.235

3,4

 

 

 

-20.360 and 20.370

20.37

3,4

 

 

 

-50.900 and 50.925

51.175

3,4

 

 

 

-101.80 and 101.85

102.35

3,4

 

 

 

< -48.925 and > 48.975

102.35

3,4

 

 

 

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

number

numeric

-101.80 to 101.85

volts

 

 

 

 

Comments Executable when initiated: No

Coupled Command: No

Calibration values on the 102.35 volt range have a special low end allowed, so that voltages much less than full scale may be used to calibrate gain on this range. Values < -48.975 and > 48.975 are accepted as legal values for calibrating the 102.35 volt range.

CALibration:VALues used for GAIN calibration are at least 85% of full scale and 10 counts away from absolute full scale. The previous table lists the gain

calibration values.

Related commands: CALibration:GAIN

*RST Condition: CALibration<chan >:VALue 1.0185

CALibration[<chan>] Subsystem

Command Reference 209

Page 209