Table 4-1. HP E1429A/B Command Quick Reference.

Subsystem

Commands

Description

 

 

 

ARM

ARM[:STARt]:COUNt <count >

Specifies the number of measurement

 

 

cycles (bursts) to occur.

 

ARM[:STARt]:DELay <period >

Delay from when the digitizer is armed to

 

 

when it enters the wait-for- trigger state.

 

ARM[:STARt][:IMMediate]

Places the digitizer in the wait-for-trigger

 

 

state, independent of the selected

 

 

ARM:STARt:SOURce.

 

ARM[:STARt]:LEVel[<chan >]:NEGative <voltage >

Selects the (negative-going) signal level

 

 

which arms the digitizer.

 

ARM[:STARt]:LEVel[<chan >]:POSitive <voltage >

Selects the (positive-going) signal level

 

 

which arms the digitizer.

 

ARM[:STARt]:SLOPe[<n >] <edge >

Selects the edge: positive, negative, or

 

 

either, which will arm the digitizer.

 

ARM[:STARt]:SOURce[<n >] <source >

Sets the digitizer arm source.

 

 

 

CALibration

CALibration[<chan >]:COUNt?

Returns a number that indicates how often

 

 

the digitizer has been calibrated.

 

CALibration[<chan >]:DATA <block_data >

Manually sets or queries the calibration

 

 

constants.

 

CALibration[<chan >]:DELay

Calibrates the A/D converter delay constant.

 

CALibration[<chan >]:GAIN [<readings >

Performs a gain calibration using the

 

[,<period >[,<flag >]]]

specified number of readings and sample

 

 

rate.

 

CALibration[<chan >]:SECure:CODE <code >

Sets the code required to disable

 

 

calibration security.

 

CALibration[<chan >]:SECure:STATe <mode > [,<code >]

Enables/disables calibration security.

 

CALibration[<chan >]:STORe

Stores the currently selected channel’s

 

 

 

 

calibration constants into non-volatile

 

CALibration[<chan >]:STORe:AUTO <mode >

memory.

 

Selects whether or not calibration constants

 

 

 

CALibration[<chan >]:VALue <number >

will be automatically stored.

 

Specifies the voltage level at the input.

 

CALibration[<chan >]:ZERO [<readings >

 

Performs a calibration of the zero offset.

 

[,<period >,[<mode >]]]

 

 

 

CONFigure

CONFigure[<chan >]:ARRay[:VOLTage][:DC] (<size>)

Configures the digitizer for <size> number

 

[,<expected value >[,<resolution >]]

of readings on the specified channel and

 

[,(@<input port>)]

input port.

 

 

 

DIAGnostic

DIAGnostic:CALibration[<chan >]:CONVerge?

Returns convergence data from the latest

 

 

CAL:ZERO or CAL:GAIN.

 

DIAGnostic:CALibration[<chan >]:GAIN:SENSitivity?

Returns the sensitivity constant used during

 

 

the last CAL:GAIN command.

 

DIAGnostic:CALibration[<chan >]:ZERO:SENSitivity?

Returns the sensitivity constant used during

 

 

the last CAL:ZERO command.

 

DIAGnostic:CHANnel[<chan >]:LABel <label >

Sets the bit pattern (label) specified on the

 

 

four least significant bits of the reading.

 

 

 

 

 

 

310 Command Quick Reference

Chapter 4

Page 310
Image 310
HP E1429A manual Command Quick Reference Chapter