HP E1429A manual Base + 4B16, Arm Control Register Power-on/Reset Settings

Models: E1429A

1 420
Download 420 pages 44.04 Kb
Page 356
Image 356

Source 1 slope: Bit 3 sets the slope of arm source 1. For all arm sources except a TTLTrg trigger line (bits 2 - 0 =001) and the HP-IB GET command or *TRG command (bits 2 - 0 =010), the slope should be set to positive (0).

Arm source 1: Bits 2 - 0 set arm trigger source 1. Arm source 2 and arm source 1 are ORed together so that an arm from either source arms the digitizer. Setting bits 2 - 0 as follows sets the arm source indicated.

0 0 0 - "Ext 1" BNC connector.

0 0 1* - TTLTrg trigger line (negative-edge triggered)

0 1 0* - HP-IB GET command or IEEE-488.2 *TRG command(negative

edge

triggered)

0 1 1

- arm when a specified input level on channel 1 is reached

1 0 0

- arm when a specified input level on channel 2 is reached

1 0 1

- ECLTrg0 trigger line

1 1 0

- ECLTrg1 trigger line

1 1 1

- arm immediate (arm source 2 must be OFF when selecting this

source)

The Arm Control Register

 

 

 

 

 

 

base + 4B16

 

The arm control register controls various digitizer arming parameters. The

 

register bits are defined below.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Address

7

6

5

 

4

3

2

1

0

 

 

 

 

 

 

 

 

 

 

 

 

base + 4B16

128

64

32

 

16

8

4

2

1

 

 

 

 

 

 

 

 

 

 

 

 

Purpose

not used

enintr0

2 speed

 

reclk/

pre-trig

thold

triginf

delay

 

 

 

 

 

 

10

 

 

 

ref

 

 

 

 

 

 

 

 

 

 

 

 

Setting

---

0 - off

0 - off

 

0 - off

0 - off

0 - off

0 - off

0 - off

 

 

 

1 - on

1 - on

 

1 - on

1 - on

1 - on

1 - on

1 - on

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Arm Control Register Power-on/Reset Settings

At power-on or following a reset, the arm control register is set to 0000 0001 or 0116.

Bit Descriptions enintr0: Enable local interrupt 0. Setting bit 6 to ’1’ enables local interrupt

0to go ’high’ if an arm trigger is received while arm source hold is set, or if the digitizer is already armed. The bit is cleared (’0’) when the digitizer is initiated.

2 speed: Setting bit 5 to ’1’ enables the dual rate sampling mode.

356 Register Programming

Appendix C

Page 356
Image 356
HP E1429A manual Base + 4B16, Arm Control Register Power-on/Reset Settings