ARM[:STARt]:LEVel[<chan>]:POSitive <voltage>

Example Setup to arm only when the signal on channel 2 falls through -0.48 volts.

ARM:SOUR1 INT2

Set arming source 1 to arm when the

 

specified level is met on channel 2

ARM:SLOP1 NEG

The signal must fall through the level set by

 

ARM:LEVel2:NEG for arming to occur

ARM:LEV2:NEG -0.48

Arm when the signal passes through the

 

-.48V level

[:STARt]:LEVel[<chan>]:POSitive <voltage>

ARM[:STARt]:LEVel[<chan>]:POSitive <voltage> selects the input voltage level which will arm a measurement cycle. The ARM:LEVel:POSitive setting is used only when either of the ARM:STARt:SOURce(s) is set to INTernal1 or INTernal2, and the corresponding ARM:STARt:SLOPe<n> setting is POSitive or EITHer. The value programmed is retained (but not used) when other sources are selected, or when the corresponding ARM:STARt:SLOPe<n> setting is NEGative.

The allowable levels depend on the measurement range as set by the

SENSe:VOLTage:RANGe command.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

1 2

none

 

 

 

 

level

numeric

-102.2418 to 102.2418

Volts

 

 

MINimum MAXimum

 

 

 

 

 

Comments Executable while initiated: No

Coupled Command: Yes. The command is coupled to the ARM[:STARt]:SLOPe, ARM[:STARt]:SOURce, and SENSe:VOLTage:RANGe commands. The level value is limited to the range limits.

ARM Subsystem

Command Reference 193

Page 193
Image 193
HP E1429A manual ARMSTARtLEVelchanPOSitive voltage