MEASure[<chan>]:ARRay[:VOLTage][:DC]?

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

1 2

none

 

 

 

 

size

numeric

1 7 to 524288

none

 

 

 

 

expected

numeric

-102.30 to 102.35

volts

value

 

DEFault

 

 

 

MINimum MAXimum

 

 

 

 

 

resolution

numeric

.00005 to .05

volts

 

 

DEFault

 

 

 

MINimum MAXimum

 

 

 

 

 

input port

numeric

13 with MEASure1

none

 

 

24 with MEASure2

 

 

 

1 and 2 are single ended

 

 

 

3 and 4 are differential

 

 

 

 

 

The maximum size parameter will be 524284 if the battery is enabled instead of 524288.

For expected value, MINimum selects the 0.100 V range and MAXimum selects either the 1.0 volt range (single ended ports) or the 100 V range (differential ports). DEFault selects 1V.

For resolution, MINimum, MAXimum, and DEFault select the same value.

Settings Maximum expected value settings per range are shown in the following table, along with the resolution associated with each range. The highlighted area shows the setting used when the expected value is not specified or DEFault is used.

MEASure[<chan>] Subsystem

Command Reference 237

Page 237
Image 237
HP E1429A manual MEASurechanARRayVOLTageDC?, Parameters