HP E1429A manual Appendix B Useful Tables

Models: E1429A

1 420
Download 420 pages 44.04 Kb
Page 339
Image 339

Table B-3. HP E1429A/B Error Messages (Cont’d)

Code

Message

Description

 

 

 

-410

Query INTERRUPTED

The digitizer was sent a command before it was

 

 

finished responding to a query command.

 

 

 

-420

Query UNTERMINATED

The controller (computer) attempts to read a query

 

 

response from the digitizer without having first sent a

 

 

complete query command.

 

 

 

-430

Query DEADLOCKED

The digitizer’s input and output buffers are full and the

 

 

digitizer cannot continue.

 

 

 

-440

Query UNTERMINATED after

Occurs when the *IDN? query is not the last query

 

indefinite response

executed in a command string.

 

 

 

1002

Cal security enabled

Calibration security must be disabled to calibrate the

 

 

digitizer, to read or write calibration data, to change

 

 

the security code, or to change the protected user

 

 

data.

 

 

 

1004

Cal write fail

Writing calibration or protected user data (*PUD) to

 

 

nonvolatile memory failed.

 

 

 

1005

Error during CAL

An error occurred during calibration. Refer to the

 

 

statement appended to this message for a description

 

 

of the error.

 

 

 

1007

Calibration security defeated

A jumper was moved to defeat calibration security.

 

 

 

1008

Error during zero cal

An error occurred during calibration of the zero offset.

 

 

Refer to the statement appended to this message for

 

 

a description of the error.

 

 

 

1009

Error during gain cal

An error occurred during gain calibration. Refer to the

 

 

statement appended to this message for a description

 

 

of the error.

 

 

 

1010

Error during linearity cal

An error occurred during linearity calibration. Refer to

 

 

the statement appended to this message for a

 

 

description of the error.

 

 

 

1015

A/D control register not responding

The serial interface register was not working properly

 

 

at power-on.

 

 

 

1016

Illegal during LBUS or VME

The command can not be executed while a VME bus

 

memory transfer

or Local bus data transfer is in progress.

 

 

 

1017

Battery too low, data may be lost

The battery does not contain sufficient charge to

 

 

maintain memory over an extended period. This error

 

 

occurs when readings are taken, when the battery is

 

 

enabled/disabled, or during the self-test.

 

 

 

1018

Battery backed data corrupt

This error is due to a low battery charge, or if the

 

 

battery is enabled after readings are in memory.

 

 

 

 

 

 

Appendix B

Useful Tables 339

Page 339
Image 339