CALibration[<chan>]:SECure:STATe

:SECure:STATe

CALibration[<chan >]:SECure:STATe <mode >,[<code >] enables or disables

calibration security. Disable the calibration security to calibrate the HP E1429, change the security code, or change the protected user data.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

chan

numeric

1 or 2

none

 

 

 

 

mode

boolean

OFF0ON1

none

 

 

 

 

code

character

1 to 12 characters

none

 

data

 

 

 

 

 

 

Comments Executable when initiated: Yes

Coupled Command: No

Either channel may be referenced, enabling/disabling is done to both channels at once with this command.

The code parameter must be present to disable the security, or it generates error -109,"Missing parameter". The value supplied must match the currently programmed security code or it generates error -224,"Illegal parameter value". A 1 second delay will then occur before the HP E1429 executes any subsequent commands.

To enable security, the code parameter is not required, but is checked if it is present. If a code is given and is incorrect, error -224, "Illegal parameter value" will be generated.

Security must be disabled to calibrate the HP E1429, or to use the *PUD command.

Related commands: CALibration:GAIN, CALibration:ZERO, CALibration:STORe, CALibration:SECure:CODE, *PUD

*RST Condition: unaffected

Example Disabling calibration security

CAL:SEC:STAT OFF,E1429

Disable security, assuming factory-set

 

security code

206 Command Reference

CALibration[<chan>] Subsystem

Page 206
Image 206
HP E1429A manual CALibrationchanSECureSTATe, CALibrationchan SECureSTATe mode ,code enables or disables