Table 4-1. HP E1429A/B Command Quick Reference (Cont’d).

Subsystem

Commands

Description

 

 

 

TRIGger

ABORt

Removes the digitizer from the

 

 

wait-for-trigger state.

 

TRIGger[:STARt]:COUNt <number >

Specifies the total number of (pre- and

 

 

post-arm) readings taken during each arm

 

 

cycle.

 

TRIGger[:STARt][:IMMediate]

Sends an immediate trigger to the digitizer.

 

 

A reading is taken if the digitizer is in the

 

 

wait-for-trigger state.

 

TRIGger[:STARt]:SOURce <source >

Sets the trigger source.

 

TRIGger[:STARt]:TIMer1 <period >

Sets the digitizer sample rate.

 

TRIGger[:STARt]:TIMer2 <period >

Sets the post-arm sample rate for dual rate

 

 

sampling.

 

 

 

VINStrument

VINStrument[:CONFigure]:LBUS:FEED <source>

Sets the Local bus data source.

 

VINStrument[:CONFigure]:LBUS:MEMory:INITiate

Places the digitizer in the wait-for-trigger

 

 

state and starts the data transfer from

 

 

digitizer memory to the Local bus.

 

VINStrument[:CONFigure]:LBUS[:MODE] <mode>

Sets the Local bus operating mode.

 

VINStrument[:CONFigure]:LBUS:RESet

Resets the digitizer’s (E1429B) Local bus

 

 

chip.

 

VINStrument[:CONFigure]:LBUS:SEND:POINts

Sets the number of readings per Local bus

 

<count>

transfer block.

 

VINStrument[:CONFigure]:LBUS:SEND:POINts:AUTO

Sets the digitizers to the Local bus

 

<mode >

interleaved transfer mode.

 

VINStrument[:CONFigure]:TEST:DATA <voltage_list>

Tests the Local bus by transmitting a list of

 

 

data.

 

VINStrument[:CONFigure]:VME:FEED <source>

Sets the VME (VXI data transfer) bus data

 

 

source.

 

VINStrument[:CONFigure]:VME:MEMory:INITiate

Places the digitizer in the wait-for-trigger

 

 

state and starts the data transfer from

 

 

digitizer memory to the VME (VXI data

 

 

transfer) bus.

 

VINStrument[:CONFigure]:VME[:MODE] <mode>

Sets the VME (VXI data transfer) bus mode.

 

VINStrument[:CONFigure]:VME:SEND:ADDRess:DATA?

Returns the A24 offset address used to

 

 

read measurement data from memory.

 

VINStrument:IDENtity?

Indicates the virtual instrument capability of

 

 

the digitizer.

 

 

 

 

 

 

Chapter 4

Command Quick Reference 313

Page 313
Image 313
HP E1429A manual VINStrument