STATus:PRESet

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

unmask

numeric or

0 through +32767

none

 

non-decimal

 

 

 

numeric

 

 

 

 

 

 

The non-decimal numeric forms are the #H, #Q, or #B formats specified by

IEEE-488.2.

Comments Executable when initiated: Yes

Coupled command: No

Related commands: STATus subsystem, *SRE, *STB?

*RST Condition: unaffected

Power-on Condition: STATUS:OPERationQUEStionable:PTRansition 32767

Example Setting the Operation register positive transition mask

STAT:OPER:PTR 64

Set event bit when wait-for-arm state is

 

entered

:PRESet

STATus:PRESet initializes the enable registers and transition masks for the Operation and Questionable Signal status registers and sets STATus:OPC:INITiate ON. For both status registers, the enable registers are set to 0, the negative transition masks are set to 0, and the positive transition masks are set to 32767.

Comments Executable when initiated: Yes

Coupled command: No

Related commands: STATus subsystem, *SRE, *STB?

*RST Condition: none

Example Presetting the STATus subsystem

STAT:PRES

Preset STATus subsystem

270 Command Reference

STATus Subsystem

Page 270
Image 270
HP E1429A manual STATusPRESet