TRIGger[:STARt]:TIMer[1]

The active edges for the various sources are as follows:

ECLtrg0, ECLtrg1 and DECLtrg active edge is the rising edge.

TTLTrg<n> and DTTLtrg active edge is the falling edge.

EXTernal1, EXTernal2, and DEXTernal active edge is the falling edge.

When using DECLtrg, DEXTernal, or DTIMer, at least one pre-arm pulse must occur after the arming signal has been received. This means that after the pre-arm count is reached and the arm is accepted, there must be another pre-arm pulse which arrives ([reference period * 3] + 60 ns) after the arming signal. This would typically be 210 ns if SENSe:ROSCillator:SOURce is INTernal. The digitizer does not sample on the additional pulse.

When TIMer is the source, the desired period must be specified by the TRIGger:STARt:TIMer1 and/or the TRIGger:STARt:TIMer2 commands. See these commands for a table of allowable values.

Related Commands: SENSe:ROSCillator:SOURce, ARM:STARt:SOURce, OUTPut subsystem, TRIGger:STARt:TIMer1, TRIGger:STARt:TIMer2, SENSe:SWEep:POINts

*RST Condition: TRIGger:STARt:SOURce TIMer

Example Setting the start trigger source

TRIG:SOUR ECLT0

A reading will be taken with each ECLT0

 

pulse

[:STARt]:TIMer[1]

TRIGger[:STARt]:TIMer[1] <period > specifies the time period between each sampling event. The time period must be a multiple of the reference oscillator period, with allowable multiples being 1,2,4,10,20,40,........1E8, 2E8, 4E8. See the

table below for the exact periods available with the internal 20 MHz reference.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

period

numeric

reference period to

seconds

 

 

reference period * 4E8

 

 

 

MINimum MAXimum

 

 

 

 

 

278 Command Reference

TRIGger Subsystem

Page 278
Image 278
HP E1429A manual TRIGgerSTARtTIMer1, Example Setting the start trigger source, Reference period to