INPut[<port>]:FILTer[:LPASs][:STATe]

:FILTer[:LPASs][:STATe]

INPut[<port>]:FILTer[:LPASs][:STATe] <mode> enables or disables the 10

MHz input filter.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

port

numeric

1234

none

 

 

 

 

mode

boolean

OFF0ON1

none

 

 

 

 

Comments ∙ Executable when initiated: Yes

∙Coupled Command: No

∙Enabling or disabling the filter while the digitizer is taking readings (initiated) may cause unexpected reading values due to relay settling during the transition.

∙Enabling the filter on an input port does not affect the input filter setting on the channel’s other port.

∙Related commands: SENSe:FUNCtion "VOLTage <port>"

∙*RST Condition: INPut<port >:FILTer:LPASs:STATe OFF

Example Enabling the 10 MHz low-pass filter

INP1:FILT:LPAS ON

Enable input filtering on input port 1

:IMPedance

INPut[<port>]:IMPedance <impedance> selects the input impedance for the

HP E1429. Either 50 or 75Ω may be selected on input ports 1 and 2, while 1 MΩ is the only setting allowed for input ports 3 and 4. Values other than 50 or 75Ω on input ports 1 and 2 will generate error -222,"Data out of range". Similarly, attempts to set input ports 3 or 4 to an impedance value other than 1 MΩ will generate error -222,"Data out of range".