Configuring the Digitizer Input

Using the A/D Shift Register

This section contains the procedures used to configure the digitizer’s input section. The configuration includes:

∙Enabling the Single Ended and Differential Inputs

∙Setting the input impedance

∙Enabling the 10 MHz filter

∙Setting the measurement range

The digitizer must be in the idle state when configuring the input section.

Each input parameter listed above is set by the digitizer’s A/D shift register. This register is accessed through the A/D status register

(base + 0316), the A/D serial register (base + 0516), and by the A/D parallel strobe register (base + 0B16) as shown in Figure C-2. The procedure for setting the configuration using the register is described in the following section.

Figure C-2. Accessing the A/D Shift Register

368 Register Programming

Appendix C

Page 368
Image 368
HP E1429A manual Using the A/D Shift Register, Figure C-2. Accessing the A/D Shift Register