MEASure[<chan>]:ARRay[:VOLTage][:DC]?

Maximum

Voltage

Resolution

Expected

Range

(Volts)

Value (V)

(Volts)

 

 

 

 

± .1

0.10235

.00005

 

 

 

± .2

0.2047

.00010

 

 

 

± .5

0.51175

.00025

 

 

 

± 1

1.0235

.0005

 

 

 

± 2

2.047

.0010

 

 

 

± 5

5.1175

.0025

 

 

 

± 10

10.235

.005

 

 

 

± 20

20.47

.010

 

 

 

± 50

51.175

.025

 

 

 

±100

102.35

.05

 

 

 

Comments ∙ Executable when initiated: No

∙Coupled Command: No

∙If no input port is given, MEASure1 defaults to (@1) and MEASure2 defaults to (@2).

∙Resolution varies by range, and is constant for each range. For any given range, specifying MINimum, MAXimum, or DEFault for resolution yields the same result.

∙It is important to note that the expected value determines the resolution and not vice versa. The digitizer always uses 12-bit resolution and a coarser resolution value has no effect. If, for some reason a coarser resolution is desired, specify a larger expected value. See the previous table for expected values and resolutions.

∙The MEASure command is equivalent to the command sequence: ABORt;CONFigure;INITiate:IMMediate;FETCh?. Because of the ABORt, VME (VXI data transfer) bus or Local bus transfers in progress will be aborted. This includes the pipelining of data.

238 Command Reference

MEASure[<chan>] Subsystem

Page 238
Image 238