MEASure[<chan>]:ARRay[:VOLTage][:DC]?

Maximum

Voltage

Resolution

Expected

Range

(Volts)

Value (V)

(Volts)

 

 

 

 

± .1

0.10235

.00005

 

 

 

± .2

0.2047

.00010

 

 

 

± .5

0.51175

.00025

 

 

 

± 1

1.0235

.0005

 

 

 

± 2

2.047

.0010

 

 

 

± 5

5.1175

.0025

 

 

 

± 10

10.235

.005

 

 

 

± 20

20.47

.010

 

 

 

± 50

51.175

.025

 

 

 

±100

102.35

.05

 

 

 

Comments Executable when initiated: No

Coupled Command: No

If no input port is given, MEASure1 defaults to (@1) and MEASure2 defaults to (@2).

Resolution varies by range, and is constant for each range. For any given range, specifying MINimum, MAXimum, or DEFault for resolution yields the same result.

It is important to note that the expected value determines the resolution and not vice versa. The digitizer always uses 12-bit resolution and a coarser resolution value has no effect. If, for some reason a coarser resolution is desired, specify a larger expected value. See the previous table for expected values and resolutions.

The MEASure command is equivalent to the command sequence: ABORt;CONFigure;INITiate:IMMediate;FETCh?. Because of the ABORt, VME (VXI data transfer) bus or Local bus transfers in progress will be aborted. This includes the pipelining of data.

238 Command Reference

MEASure[<chan>] Subsystem

Page 238
Image 238
HP E1429A manual Maximum Voltage Resolution