[SENSe[<chan>]]:VOLTage[:DC] :RANGe

Range

Measurement

Resolution

Allowable

Setting

Range (V)

(Volts)

 

(V)

 

 

Ports

 

 

 

 

0.10235

-.10230 to .10235

.00005

1,2,3,4

 

 

 

 

0.2047

-.2046 to .2047

.00010

1,2,3,4

 

 

 

 

0.51175

-.5115 to .51175

.00025

1,2,3,4

 

 

 

 

1.0235

-1.0230 to 1.0235

.0005

1,2,3,4

 

 

 

 

2.047

-2.046 to 2.047

.0010

3,4

 

 

 

 

5.1175

-5.115 to 5.1175

.0025

3,4

 

 

 

 

10.235

-10.230 to 10.235

.005

3,4

 

 

 

 

20.470

-20.460 to 20.470

.010

3,4

 

 

 

 

51.175

-51.150 to 51.175

.025

3,4

 

 

 

 

102.35

-102.30 to 102.35

.05

3,4

 

 

 

 

Comments Executable when initiated: Yes

Coupled Command: Yes. The command is coupled to ARM[:STARt]:LEVel[<chan >]:NEGative, ARM[:STARt]:LEVel[<chan >]:POSitive, and SENSe[<chan >]:FUNCtion.

Though the range setting may be changed while the HP E1429 is taking readings (INITiated), it will take at least 3 ms for the relay to settle. Also, it is up to the user to determine where in the data archive the new range setting was switched in, as the HP E1429 does not stamp the data in any way. If the data format is not PACKed, then the HP E1429 will use the resolution associated with the final range setting to convert all readings into voltage values when FETCh? is executed.

MAXimum values for the range setting will depend on the current SENSe:FUNCtion setting for that channel. For single-ended ports (1 and 2), MAXimum is 1.0235, and for differential ports (3 and 4) MAXimum is 102.35.

*RST Condition: SENSe1:VOLTage:DC:RANGe 1.0235, and SENSe2:VOLTage:DC:RANGe 1.0235

[SENSe[<chan>]]:VOLTage[:DC] Subsystem

Command Reference 263